ECG: Lead mode: 5-leads (RA, LA, RL, LL, V) Lead method: I, II, III, aVR, aVL, aVF and V Waves: 9/7 channels Gain: ×0,5; ×1; ×2; ×4 Scan speed: 12,5, 25, 50 mm/s Modes: DIA 0,05-75 Hz; MON 0,5-35 Hz; OPE 1-25 Hz Protected against defibrillation shocks
HEART RATE (HR): Range: 0-300 bpm Accuracy: ±1 bpm or 2% Resolution: 1 bpm ECG calibration: 1 mV (peak value) ST segment analysis: -1,0 mV to +1,0 mV Accuracy: ±0,02 mV or ±5%
NIBP: Method: Oscillometric Operation mode: Manual, Auto Measurement interval (Auto): 1-60 min PR range: 0-300 bpm Units: mmHg, kPa Measurement: Systolic, Diastolic, Mean Suitable for Adults, Children, Neonates Overpressure protection
RESPIRATION: Method: RA-LL impedance Range: 1-99 rpm Resolution: 1 rpm Accuracy: ±2 rpm
TEMPERATURE: Channel: 2 Range: 0-45 °C Resolution: 0,1 °C Accuracy: ±0,2 °C
SpO2: Range: 0-100% Resolution: 1% Accuracy: 80%-100%; ±2% 60%-79%; ±3% Pulse range: 0-300 bpm
| | 1 Power cord 1 Rechargeable Li-ion-Battery 1 User's manual 1 3- or 5-lead ECG cable (snap or clip) 1 Blood pressure cuff with extension hose 1 Resusable SpO2 finger probe, adult 1 Temperature probe 1 Warranty card |